  1. Pubblicazioni

CRITICAL APPRAISAL OF VOLUMETRIC MODULATED ARC THERAPY IN STEREOTACTIC BODY RADIATION THERAPY FOR METASTASES TO ABDOMINAL LYMPH NODES

Articolo
Data di Pubblicazione:
2009
Abstract:
Purpose: A planning study was performed comparing volumetric modulated arcs, RapidArc (RA), fixed beam
IMRT (IM), and conformal radiotherapy (CRT) with multiple static fields or short conformal arcs in a series of
patients treated with hypofractionated stereotactic body radiation therapy (SBRT) for solitary or oligo-metastases
from different tumors to abdominal lymph nodes.
Methods and Materials: Fourteen patients were included in the study. Dose prescription was set to 45 Gy (mean
dose to clinical target volume [CTV]) in six fractions of 7.5 Gy. Objectives for CTV and planning target volume
(PTV) were as follows: Dosemin >95%, Dosemax <107%. For organs at risk the following objectives were used: Maximum
dose to spine <18 Gy; V15Gy <35% for both kidneys, V36Gy <1% for duodenum, V36Gy <3% for stomach and
small bowel, V15Gy <(total liver volume 700 cm3) for liver. Dose–volume histograms were evaluated to assess plan
quality.
Results: Planning objectives on CTVand PTV were achieved by all techniques. Use of RA improved PTV coverage
(V95% = 90.2% ± 5.2% for RA compared with 82.5% ± 9.6% and 84.5% ± 8.2% for CRT and IM, respectively).
Most planning objectives for organs at risk were met by all techniques except for the duodenum, small bowel,
and stomach, in which the CRT plans exceeded the dose/volume constraints in some patients. The MU/fraction
values were as follows: 2186 ± 211 for RA, 2583 ± 699 for IM, and 1554 ± 153 for CRT. Effective treatment
time resulted as follows: 3.7 ± 0.4 min for RA, 10.6 ± 1.2 min for IM, and 6.3 ± 0.5 min for CRT.
Conclusions: Delivery of SBRT by RA showed improvements in conformal avoidance with respect to standard conformal
irradiation. Delivery parameters confirmed logistical advantages of RA, particularly compared with
IM.
Tipologia CRIS:
1.1 Articolo in rivista
Keywords:
Stereotactic body radiation therapy; Volumetric modulated arc therapy; IMRT
Elenco autori:
Bignardi, M; Cozzi, L; Fogliata, A; Lattuada, P; Mancosu, P; Navarria, P; Urso, G; Vigorito, S; Scorsetti, M
Autori di Ateneo:
Mancosu Pietro
Scorsetti Marta
Link alla scheda completa:
https://iris.hunimed.eu/handle/11699/3953
Pubblicato in:
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS